Validation of dynamic three-dimensional whole heart magnetic resonance myocardial perfusion imaging at 3.0 Tesla against fractional flow reserve for the detection of flow-limiting coronary heart disease
Journal of Cardiovascular Magnetic Resonance volume 14, Article number: O92 (2012)
Summary
We demonstrate the feasibility of 3D myocardial perfusion CMR at 3 Tesla against fractional flow reserve (FFR) in 53 patients for the detection of flow-limiting coronary artery disease and show good agreement between the techniques. This technique shows excellent diagnostic sensitivity and specificity and may offer an alternative method of detecting ischaemia for the purpose of guiding revascularisation and risk stratification.
Background
Three-dimensional (3D) myocardial perfusion cardiovascular magnetic resonance (CMR) has recently been proposed to overcome the limited spatial coverage of conventional perfusion CMR methods1. The method has shown good diagnostic accuracy for the detection of coronary artery disease determined by quantitative coronary angiography (QCA)2. However the relationship between the severity of a coronary stenosis on QCA and its functional significance is variable. Pressure wire-derived fractional flow reserve (FFR) <0.75 correlates closely with objective evidence of reversible ischemia and it has been demonstrated that ischaemia-guided PCI confers a prognostic benefit.
Aim
To determine the diagnostic accuracy of whole heart 3D myocardial perfusion CMR against invasively determined FFR.
Methods
Fifty-five patients referred for angiography underwent rest and adenosine stress 3D myocardial perfusion CMR at 3Tesla (3D turbo gradient echo, flip angle 15, TR 2.0ms/TE 1.0ms, 12 slices of 5mm thickness, in-plane resolution 2.3x2.3mm2, 10 fold k-space and time k-t broad linear speed up technique acceleration with k-t principal component analysis). Perfusion was scored visually on a patient and coronary territory basis with a score from 0-3. Ischaemic burden was calculated by quantitative segmentation of the volume of hypoenhancement. The FFR was measured in vessels with >50% severity stenosis. Fractional flow reserve <0.75 was considered hemodynamically significant.
Results
Two patients were excluded (one claustrophobia, on poor image quality). From the remaining fifty-three patients and 159 coronary vessels, 64 underwent pressure wire assessment and 39 had an FFR<0.75 (Figure1). Sensitivity, specificity and diagnostic accuracy of CMR analysis per patient was 90%, 91% and 91%, respectively for the detection of significant coronary artery disease. By coronary territory the values were 79%, 92% and 88%.
Conclusions
3D CMR stress perfusion can detect functionally significant coronary artery disease with excellent sensitivity, specificity and predictive values when compared with FFR.
3D CMR perfusion imaging may offer an alternative method of detecting ischaemia for the purpose of guiding revascularisation and risk stratification.
Funding
SP is funded by British Heart Foundation fellowship FS/10/62/28409.
SP/EN receives research grant support from Philips Healthcare.
References
Vitanis V, Manka R, Giese D, Pedersen H, Plein S, Boesiger P, Kozerke S: High resolution three-dimensional cardiac perfusion imaging using compartment-based k-t principal component analysis. Magn Reson Med. 2011, 65 (2): 575-87. 10.1002/mrm.22620. doi: 10.1002/mrm.22620. Epub 2010 Oct 6
Manka R, Jahnke C, Kozerke S, Vitanis V, Crelier G, Gebker R, Schnackenburg B, Boesiger P, Fleck E, Paetsch I: Dynamic 3-dimensional stress cardiac magnetic resonance perfusion imaging detection of coronary artery disease and volumetry of myocardial hypoenhancement before and after coronary stenting. J Am Coll Cardiol. 2011, 57 (4): 437-44. 10.1016/j.jacc.2010.05.067.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Jogiya, R., Morton, G., De Silva, K. et al. Validation of dynamic three-dimensional whole heart magnetic resonance myocardial perfusion imaging at 3.0 Tesla against fractional flow reserve for the detection of flow-limiting coronary heart disease. J Cardiovasc Magn Reson 14 (Suppl 1), O92 (2012). https://doi.org/10.1186/1532-429X-14-S1-O92
Published:
DOI: https://doi.org/10.1186/1532-429X-14-S1-O92