Figure 2

From: Phase contrast MRI with flow compensation view sharing (FCVS)

Figure 2

(a) The FC signal phase as a function of the cardiac frames for a randomly selected pixel within the CCA (mean/± SD: -2.51/± 0.065). (b) Peak velocity waveforms from the standard FCFE PC-MRI (gray curve) with 17 ms-temporal-resolution, FCVS (blue curve) with 17 ms-temporal-resolution, standard FCFE PC-MRI (red curve) with 34 ms-temporal-resolution. The FCVS results are highly correlated with the measurements from standard FCFE PC-MRI at the same temporal resolution but FCVS only requires 50% of the acquisition time. The standard FCFE PC-MRI fails to capture the peak velocity at approximately 75 ms or the transient dip at 320 ms when its temporal resolution is halved to match the total acquisition time of FCVS. (c) The Bland-Altman plot of total volumetric flow measurements between standard FCFE PC-MRI and FCVS with two different temporal resolutions (17 ms and 34 ms) in the left and right CCA in six volunteers for a total 24 flow measurement.