Table 4 Intrareader variability for all contours combined

From: Repeatability and reproducibility of various 4D Flow MRI postprocessing software programs in a multi-software and multi-vendor cross-over comparison study

 

SW-A

SW-B

SW-C

SW-D

Stroke volume

Absolute error [ml]

0 ± 6

 − 1 ± 18

2 ± 6

0 ± 12

Relative error [%]

1 ± 7

 − 2 ± 26

3 ± 8

0 ± 16

ICC

0.967***

0.648*

0.964***

0.790**

Peak flow

Absolute error [ml/s]

4 ± 27

 − 3 ± 89

26 ± 92

5 ± 41

Relative error [%]

1 ± 7

 − 1 ± 27

8 ± 27

1 ± 11

ICC

0.969***

0.644*

0.663*

0.917***

Peak maximum velocity

Absolute error [cm/s]

 − 1 ± 11

 − 1 ± 28

5 ± 26

 − 2 ±  − 20

Relative error [%]

 − 1 ± 8

 − 1 ± 21

4 ± 19

 − 1 ± 13

ICC

0.883**

0.219

0.348

0.746*

Peak average velocity

Absolute error [cm/s]

 − 1 ± 10

 − 2 ± 17

9 ± 10

3 ± 7

Relative error [%]

 − 1 ± 14

 − 2 ± 21

12 ± 13

4 ± 8

ICC

0.809**

0.394

0.541*

0.839**

Area

Absolute error [mm2]

15 ± 97

13 ± 190

 − 80 ± 143

 − 21 ± 87

Relative error [%]

3 ± 15

2 ± 29

 − 12 ± 19

 − 4 ± 15

ICC

0.919***

0.705*

0.788**

0.915***

Peak WSS per segment

Absolute error [mPa]

8 ± 133

737 ± 847

 − 137 ± 614

Relative error [%]

1 ± 9

44 ± 50

 − 5 ± 29

ICC

0.921***

0.068

0.677*

Peak WSS per contour

Absolute error [mPa]

14 ± 121

 − 32 ± 54

567 ± 563

 − 12 ± 156

Relative error [%]

2 ± 11

 − 7 ± 12

60 ± 56

 − 1 ± 22

ICC

0.921***

0.944***

0.068

0.706*

  1. Absolute and relative error are given as mean ± standard deviation. Values of bias and standard deviation yielding to smallest limits of agreement within one row are marked as bold indicating that they were used to calculate equivalence limits. Asterisks indicate ***excellent, **good, and *moderate intraclass correlation (ICC). No asterisk indicates poor agreement. WSS, wall shear stress