Table 1 4D flow acquisition parameters for large vessels and whole heart

From: 4D Flow cardiovascular magnetic resonance consensus statement: 2023 update

Acquisition parameter

Aim for

Do not exceed

Reason

Limiting factor

Acquired spatial resolution

Adults whole heart

2.5 mm3 [2]

3 mm3

Accuracy

Scan time

Adults’ vessels

2 mm3

2.5 mm3

Accuracy

Scan time

Pediatric whole heart

2 mm3 [53, 79, 82]

2.5 mm3 [82, 167]

Accuracy

Scan time

Pediatric vessels

1.5 mm3

2 mm3

Accuracy

Scan time, VNR

Neonates

0.75–1 mm3 [54]

1.5 mm3

Accuracy

Scan time, VNR

Acquired temporal resolution

30 ms

50 ms (can be higher if aiming for visualization only)

Accuracy

Scan time

Velocity encoding limit (VENC)

Maximum expected velocity

10% higher than maximum expected velocity, do not exceed 25%

VNR, to avoid aliasing

Scan time, VNR

ECG gating

Retrospective

 

Complete ECG cycle included

Reconstruction

Respiratory navigation

Optional

 

Accuracy

Scan time

Contrast agent

Consider in neonates, dissection patients or other challenging cases

 

Improved contrast

Contrast administration contraindications

Flip angle

7° non-contrast

12–25° with contrast

25°

SNR

Contrast vs SNR