From: 4D Flow cardiovascular magnetic resonance consensus statement: 2023 update
Acquisition parameter | Aim for | Do not exceed | Reason | Limiting factor |
---|---|---|---|---|
Acquired spatial resolution | ||||
Adults whole heart | 2.5Â mm3 [2] | 3Â mm3 | Accuracy | Scan time |
Adults’ vessels | 2 mm3 | 2.5 mm3 | Accuracy | Scan time |
Pediatric whole heart | Accuracy | Scan time | ||
Pediatric vessels | 1.5Â mm3 | 2Â mm3 | Accuracy | Scan time, VNR |
Neonates | 0.75–1 mm3 [54] | 1.5 mm3 | Accuracy | Scan time, VNR |
Acquired temporal resolution | 30Â ms | 50Â ms (can be higher if aiming for visualization only) | Accuracy | Scan time |
Velocity encoding limit (VENC) | Maximum expected velocity | 10% higher than maximum expected velocity, do not exceed 25% | VNR, to avoid aliasing | Scan time, VNR |
ECG gating | Retrospective | Â | Complete ECG cycle included | Reconstruction |
Respiratory navigation | Optional | Â | Accuracy | Scan time |
Contrast agent | Consider in neonates, dissection patients or other challenging cases | Â | Improved contrast | Contrast administration contraindications |
Flip angle | 7° non-contrast 12–25° with contrast | 25° | SNR | Contrast vs SNR |