Skip to main content

Table 5 A comparative analysis of sensitivity, specificity, and AUC between WHCA with and without vasodilators

From: Diagnostic accuracy of whole heart coronary magnetic resonance angiography: a systematic review and meta-analysis

 

1.5 T WHCA with vasodilator (N = 12 studies)

1.5 T WHCA without vasodilator (N = 9 studies)

P-value

Patient-based analysis

 Sensitivity

0.86 (0.80–0.90)

0.88 (0.75–0.95)

0.85

 Specificity

0.76 (0.69–0.82)

0.65 (0.58–0.82)

0.11

 AUC

0.88

0.86

N/A

Vessel-based analysis

 Sensitivity

0.81 (0.74–0.87)

0.91 (0.83–0.96)

0.03

 Specificity

0.87 (0.79–0.92)

0.73 (0.50–0.88)

0.17

 AUC

0.9

0.92

N/A

Segment-based analysis

 Sensitivity

0.84 (0.79–0.87)

0.80 (0.65–0.89)

0.54

 Specificity

0.92 (0.87–0.96)

0.93 (0.85–0.96)

0.78

 AUC

0.85

0.93

N/A

  1. Data are weighted mean ± 95% confidence interval
  2. AUC area under the curve, WHCA whole heart coronary magnetic resonance angiography