Table 1 Scanner and sequence parameters for T1 and T2 mapping acquisitions

From: Multi-site comparison of parametric T1 and T2 mapping: healthy travelling volunteers in the Berlin research network for cardiovascular magnetic resonance (BER-CMR)

Parameter

Site I

Site II

Site III

Field strength (T)

3

3

3

Vendor

Siemens

Siemens

Siemens

Model

SkyraFIT

PrismaFIT

PrismaFIT

RF-coil

18-channel

32-channel

18-channel

Bore size (cm)

70

60

60

Bore size (cm)

70

60

60

T1 scan parameters

   

Sequence

5-3-3 MOLLI

5-3-3 MOLLI

5-3-3 MOLLI

Slice thickness (mm)

6.0

6.0

6.0

Voxel size (mm × mm)

1.4 × 1.4

1.4 × 1.4

1.4 × 1.4

Field of view (mm)

360

360

360

Echo time (ms)

1.12

1.12

1.12

Repetition time (ms)

3.9

3.9

3.9

Flip angle (o)

35

35

35

Bandwith (Hz/Px)

1085

1085

1085

Acceleration factor

2

2

2

T2 scan parameters

   

Sequence

FLASH

FLASH

FLASH

Slice thickness (mm)

6.0

6.0

6.0

Voxel size (mm × mm)

1.9 × 1.9

1.9 × 1.9

1.9 × 1.9

Field of view (mm)

360

360

360

Echo time (ms)

1.32

1.32

1.32

Repetition time (ms)

3.6

3.6

3.6

Flip angle (degrees)

12

12

12

Bandwidth (Hz/Px)

1184

1184

1184

Acceleration factor

2

2

2

  1. Modified Look-Locker inversion recovery sequence (MOLLI); fast low-angle shot (FLASH); steady-state free-precession (SSFP)