Skip to main content

Table 5 Intra-observer and inter-observer reproducibility of T1 and T2 values

From: Reference ranges of myocardial T1 and T2 mapping in healthy Chinese adults: a multicenter 3T cardiovascular magnetic resonance study

 

Bias

SD of bias

95% LOA

CoV (%)

ICC

p value

Intra-observer

      

T1 base

4.63

7.19

(− 18.73 to 9.46)

2.6

0.962

< 0.001

T1 middle

− 5.19

6.14

(− 17.23 to 6.85)

2.6

0.970

< 0.001

T1 apex

− 4.23

8.19

(− 20.28 to 11.83)

3.5

0.976

< 0.001

T1 global

− 4.54

4.14

(− 12.65 to 3.57)

2.4

0.978

< 0.001

T2 base

− 0.34

0.80

(− 1.90 to 1.20)

6.2

0.926

< 0.001

T2 middle

− 0.58

0.94

(− 2.42 to 1.27)

6.6

0.892

< 0.001

T2 apex

− 0.38

0.86

(− 2.06 to 1.30)

7.9

0.948

< 0.001

T2 global

− 0.45

0.72

(− 1.87 to 0.97)

6.3

0.930

< 0.001

Inter-observer

      

T1 base

− 5.58

9.23

(− 23.68 to 12.52)

2.6

0.939

< 0.001

T1 middle

− 4.92

7.62

(− 19.84 to 10.01)

2.7

0.967

< 0.001

T1 apex

− 3.72

9.20

(− 21.75 to 14.32)

3.6

0.973

< 0.001

T1 global

− 4.58

5.29

(− 14.95 to 5.78)

2.5

0.972

< 0.001

T2 base

− 0.35

0.83

(− 1.98 to 1.28)

6.1

0.917

< 0.001

T2 middle

− 0.65

1.03

(− 2.67 to 1.38)

6.6

0.869

< 0.001

T2 apex

− 0.37

0.90

(− 2.14 to 1.40)

7.9

0.943

< 0.001

T2 global

− 0.47

0.78

(− 2.00 to 1.06)

6.2

0.918

< 0.001

  1. LOA  limit of agreement, CoV coefficient of variation, ICC intra-class correlation coefficient, SD standard deviation