Figure 1 | Journal of Cardiovascular Magnetic Resonance

Figure 1

From: Three dimensional first-pass myocardial perfusion imaging at 3T: feasibility study

Figure 1

Pulse sequence diagram and sampling pattern for 3D myocardial perfusion imaging. The pulse sequence consists of a 90° BIR-4 adiabatic RF pulse for myocardial saturation, followed by a saturation recovery time and undersampled 3DFT GRE readout. 3D k-space was undersampled by factor of three and two in ky and in kz, respectively, resulting in a net undersampling factor of six. The k-space sampling locations were shifted in a cyclic pattern such that coil sensitivity map could be obtained from the data in six neighboring R-R intervals. SENSE reconstruction of undersampled data from the kthcardiac cycle utilized coil sensitivity maps derived from the (k-2)thto (k+3)thcardiac cycles.