Skip to main content

Table 1 SNR comparison. The combination of a 32-channel cardiac array and Karhunen-Loeve image filter is utilized to maintain SNR in highly accelerated real-time cine imaging.

From: 1110 Improved real-time cine with the use of a 32-channel cardiac array and Karhunen-Loeve Transform filter

 

12-channel unfiltered

32-channel unfiltered

p value 2-channel vs. 32-channel unfiltered

32-channel KLT-filtered

p value 32-channel unfiltered vs. KLT filtered

No TSENSE

97.8 ± 20.7

90.6 ± 26.3

0.4775

n/a

n/a

TSENSE Rate 3

31.9 ± 2.4

28.6 ± 4.3

0.1151

61.9 ± 10.3

0.0011

TSENSE Rate 4

20.4 ± 3.7

22.2 ± 3.8

0.0605

48.8 ± 6.5

< 0.001

TSENSE Rate 5

10.9 ± 3.3

15.2 ± 4.7

0.0556

38.6 ± 9.3

< 0.001