Table 1 Largest ROI mean offset (cm/s) within 50 mm of Isocentre, for aortic slice, MPA slice (HF phase-encoding) and MPA slice (LR phase-encoding). The four rows per slice are from the four sites using each scanner type. The column order is not specified, i.e. scanner types are not identified. (< ± 0.3 cm/s stdev total error as described above)

From: Evidence across CMR sites and systems of background velocity offset errors requiring correction before accurate measurement of regurgitant and shunt flow

cm/s

Site

Scanner type 1

Scanner type 2

Scanner type 3

Aorta

1

2.5

1.4

1.7

 

2

0.7

2.5

0.9

 

3

0.9

2.9

1.4

 

4

1.9

1.6

0.7

MPA (HF phase-enc)

1

0.8

3.9

1.0

 

2

1.1

3.4

1.3

 

3

0.8

4.9

1.2

 

4

1.0

3.9

0.8

MPA (LR phase-enc)

1

1.2

3.3

1.5

 

2

1.0

5.3

1.6

 

3

0.8

5.0

1.8

 

4

0.4

3.6

0.6

Maximum per site

1

2.5

3.9

1.7

 

2

1.1

5.3

1.6

 

3

0.9

5.0

1.8

 

4

1.9

3.9

0.8