Novel technique of strain assessment utilizing feature tracking in nontagged SSFP images: validation with tagged strain analysis

Introduction

Recent cardiac MRI (CMR) studies have demonstrated decline in left ventricular peak circumferential strain (εcc) despite normal ejection fraction in Duchenne muscular dystrophy (DMD) patients. However, these analyses used CMR tagging, a technique limited by tag fading and complicated analysis requirements. Feature tracking software has recently been developed for analysis of εcc from non-tagged standard steady-state free precession (SSFP) cine CMR images.

Purpose

The purpose of this study was to compare Mid-LV slice εcc by feature tracking of SSFP cine CMR images to HARP analysis of tagged images.

Methods

εcc was assessed from CMR SSFP short-axis cine stack images and cine myocardial tagged image data of 54 DMD patients and 6 aged-matched control subjects, utilizing both tagged analysis (via HARP® software, Diagnosoft Inc) and SSFP feature tracking (via DIOGENES® software, TomTec Inc) methods. Analyses were performed on identical location mid-papillary LV slices (both tagged and standard SSFP). Average εcc was tabulated and compared via Spearmen rank correlation and Bland-Altman comparison of methods.

Results

Feature tracking εcc analysis correlated favorably with tagged HARP analysis (fig 1a). In addition, the techniques do not demonstrate systematic over or underestimation of one another, though the limits of agreement are relatively wide (fig 1b).

figure 1

Figure 1

Conclusion

CMR feature tracking is a feasible method for assessment of εcc. Further study on larger subject groups is warranted to determine efficacy and accuracy of feature tracking versus tagged analysis.

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution 2.0 International License (https://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Hor, K.N., Wash, E., Fleck, R.J. et al. Novel technique of strain assessment utilizing feature tracking in nontagged SSFP images: validation with tagged strain analysis. J Cardiovasc Magn Reson 11 (Suppl 1), P33 (2009). https://doi.org/10.1186/1532-429X-11-S1-P33