From: Baseline correction of phase-contrast images in congenital cardiovascular magnetic resonance

Figure 3

Phase Contrast Flow vs. Time Curves. Flow versus time curves from one cardiac cycle for the main pulmonary artery without (blue) and with (red) phantom correction. Flow volumes were calculated by integrating the area under the curves. With phantom correction, the flow in the main pulmonary artery changed from 5.1 to 2.4 L/min/m2, and the regurgitant fraction changed from 0% to 19%.