Figure 8 | Journal of Cardiovascular Magnetic Resonance

Figure 8

From: Real-time cardiovascular magnetic resonance at high temporal resolution: radial FLASH with nonlinear inverse reconstruction

Figure 8

Interleaved dual-slice real-time radial FLASH CMR. T1-weighted images (spoiled multi-slice FLASH, TR = 4.0 ms for two sections, TE = 2.0 ms, flip angle 10°, 11 spokes, 44 ms total acquisition time) of two simultaneously acquired short-axis views (a) in the apical segment and (b) 20 mm above in the mid-cavity segment (2.0 mm resolution, 8 mm section thickness). The images cover a post-systolic expansion phase by every second frame (88 ms apart). They were selected from respective real-time CMR movies reconstructed by nonlinear inversion and zoomed by a factor of 1.5.