Figure 3 | Journal of Cardiovascular Magnetic Resonance

Figure 3

From: Shortened Modified Look-Locker Inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 1.5 and 3 T within a 9 heartbeat breathhold

Figure 3

Relationship between T1 measurements for ShMOLLI (top row) and MOLLI (bottom row) and the corresponding reference T1 (T1ref) depending on selected simulated heart rates (HR). Diagonal line represents the ideal identity line. Simulation (A&C) and Phantom (B&D) measurements in 3T and 1.5T (dashed lines) overlap and are in close agreement with simulation results. NOTE. The lines are offset by 15 ms horizontally for each HR value to reduce the overlap between lines.