From: The use of parallel imaging techniques for the measurement of T2* decay
Sequence | Routine acquisition (mean ± SD) ms | With parallel imaging (mean ± SD) ms | P value | Coefficient of variation (%) |
---|---|---|---|---|
Cardiac T2* bright blood | 29.5 ± 12.6 | 29.8 ± 12.9 | 0.37 | 8.16 |
Cardiac T2* dark blood | 26.02 ± 10.4 | 26.28 ± 10.6 | 0.14 | 5.31 |
Hepatic T2* | 5.7 ± 4.8 | 5.8 ± 4.6 | 0.20 | 4.81 |