From: Direct cooling of the catheter tip increases safety for CMR-guided electrophysiological procedures

Figure 4

Ablation generator induced temperature rises. Temperature distribution measured at increasing tissue depth from the catheter tip. Different power levels were examined. The upper figure shows the temperature development at a low contact force. The lower figure shows a similar development of temperature that was measured at a high contact force. The distribution of heat is not significantly influenced by the contact pressure. Additionally, the temperature decline due to increasing distance from the catheter tip remains unchanged.