Skip to main content

Table 4 Patients with univentricular heart physiology: Mean values and agreement of 2D and 4D velocity acquisition measurements

From: Systemic-to-pulmonary collateral flow in patients with palliated univentricular heart physiology: measurement using cardiovascular magnetic resonance 4D velocity acquisition

 

BCPC

Fontan

Univentricular heart physiology

 

2D velocity acquisition

4D velocity acquisition

2D velocity acquisition

4D velocity acquisition

Bland-Altman

Pearson

difference

Correlation

2D - 4D velocity acquisition

2D - 4D velocity acquisition

SVC

1.77±0.61

1.62±0.61

0.93±0.34

0.92±0.35

0.07 ± 0.04

0.96 *

IVC

1.79±1.01

1.74±1.05

1.80±0.63

1.64±0.55

0.11 ± 0.01

0.93 *

RPA

1.06±0.37

1.01±0.36

1.43±0.51

1.47±0.59

0.01 ± 0.05

0.91*

LPA

0.79±0.37

0.77±0.44

1.12±0.28

1.10±0.33

0.02 ± 0.05

0.94*

AO

3.61±1.21

3.57±1.16

3.08±0.70

3.04±0.56

0.04 ± 0.05

0.94*

SPCF

0.59±0.52

0.79±0.45

0.41±0.46

0.62±0.82

−0.02 ± 0.18

0.73*

Qp:Qs

(0.43±0.23):1

(0.42±0.23):1

(0.82±0.29):1

(0.84±0.26):1

(0.02 ± 0.18):1

0.82*

  1. Values expressed as l/min/m2. *statistically significant correlation; 2D, two-dimensional; 4D, four-dimensional, AO, aorta; BCPC, bilateral cavopulmonary connection; CMR, cardiovascular magnetic resonance; IVC, inferior vena cava; LPA, left pulmonary artery; Qp, pulmonary blood flow; Qs, systemic blood flow; RPA, right pulmonary artery; SPCF, systemic-to-pulmonary collateral flow (SPCF = AO-SVC-IVC); SVC, superior vena cava.