Figure 8 | Journal of Cardiovascular Magnetic Resonance

Figure 8

From: Cardiovascular magnetic resonance physics for clinicians: part II

Figure 8

The relationship between signal intensity and concentration. Signal intensity values over a range of concentrations for a spoiled gradient echo pulse sequence. The two dashed curves show the separate dependencies of the signal behaviour for T1 or T2 alone. The solid line shows the combined effect of T1 or T2 on signal intensity. At low concentrations the effect of T1 shortening is dominant, while at higher concentrations the T2 shortening effect becomes the dominant factor. A series of samples imaged with increasing percentage concentrations of Gadolinium are shown underneath the plot as a visual demonstration of the effect.