Skip to main content

Table 1 MRI sequences for brain scan

From: Monocenter feasibility study of the MRI compatibility of the Evia pacemaker in combination with Safio S pacemaker lead

 

Sequences for Philips MRI Scanner

 

Body Region: Head, Patient Position: Supine, Patient Entry: Head First, Landmark on Eyes

 

3 plane localizer

Reference Scan

AX SE T1

AX FSE T2

T2 FLAIR

Diffusion

3D TOF MT

CE-MRA

Perfusion

 

Landmark on Eyes

Landmark on Eyes

Landmark on Eyes

Landmark on Eyes

Landmark on Eyes

Landmark on Eyes

Landmark on Eyes

Landmark on Eyes

Landmark on Eyes

Name Anatomy Philips

Head CNS-Brain

Ref Scans

Head CNS-Brain

Head CNS-Brain

Head CNS-Brain

Head CNS-Function

Head CNS-Angio

Head CNS-CE-Angio

Head CNS-Function

Sequence Name Philips

survey

RefScan

T1W_SE

T2W_TSE

T2W_FLAIR

DWI_E

S3DI_MC_HR

3D_DYN_AVM

sPRESTO

plane

ax/cor/sag

-

ax

ax

ax

ax

ax

sag

ax

Scan time

0:17

0:51

02:39

02:13

03:18

00:36

02:52

00:51

0:50

stacks

3

2

1

1

1

1

1

1

1

slices

3 each

82

18

18

16

18

100

10

30

Scan technique

T1-TFE

SENSE

SE

SE

IR

SE

FFE

FFE

FFE

Fast imaging mode

TFE

-

MS

TSE

TSE

EPI

none

none

EPI

(Turbo/EPI) Factor

42

-

1

19

23

89

-

-

19

TE [ms]

5.2

0.95

15

100

120

104

6.9

1.27

shortest

TR [ms]

15

8

483

3607

6000

3034

24

5

22

TI [ms]

-

-

-

-

2000

-

-

-

-

Flip angle [deg]

20

7

140

90

100

90

20

35

7

FOV [mm]

250

-

230

230

230

230

200

260

220

Scan [%]

74

82

79

74

68

79.6

57.4

100

55.8

Scan matrix

256 x 126

64 x 52

256 x 163

384 x 228

208 x 115

112 x 89

332 x 190

160 x 144

64 x 29

Slice thickness [mm]

acq/rec

10

-

5

5

6

5

1 / 0.5

16 / 8

3.5

Slice gap

0

 

1

1

2

1

-

-

-

NSA / NEX

1

3

2

3

2

1

1

1

1

Bandwidth [Hz/pixel]

186.9

2540.7

109.3

224.0

152.9

1834

108.8

541.1

60.3