Table 5 Lead Measurements and Battery status

From: Monocenter feasibility study of the MRI compatibility of the Evia pacemaker in combination with Safio S pacemaker lead

 

Follow-up

Parameter

1stFU (n = 30)

2ndFU (n = 30)

3rdFU (n = 29)

4thFU (n = 29)

p = *

Time from MRI (months)

-

-

1.1 ± 0.3

3.0 ± 0.3

 

RA sensing (mV)

3.2 ± 2.1

3.2 ± 2.3

3.2 ± 2.1

3.1 ± 2.0

n.s.

RA pacing threshold (V@0,4 ms)

0.68 ± 0.18

0.67 ± 0.16

0.71 ± 0.24

0.73 ± 0.15

n.s.

RA pacing impedance (Ohms)

507 ± 55

500 ± 46

520 ± 50

495 ± 59

n.s.

RV sensing (mV)

15.0 ± 6.0

15.0 ± 6.0

14.9 ± 6.5

14.7 ± 6.3

n.s.

RV pacing threshold (V@0,4 ms)

0.78 ± 0.22

0.79 ± 0.20

0.78 ± 0.22

0.82 ± 0.24

n.s.

RV pacing impedance (Ohms)

608 ± 54

599 ± 52+

607 ± 47

597 ± 57

n.s.

Battery status (%)

100 ± 0

100 ± 0

100 ± 0

99.7 ± 1.3

n.s.

Value changes (% [median]; compared with pre MRI FU)

     

RA sensing

 

−2 ± 20 [0]

+6 ± 43 [0]

+15 ± 57 [8]

 

RA PTH

 

+2 ± 23 [0]

+2 ± 30 [0]

+9 ± 7 [0]

 

RA Pimp

 

−1 ± 4 [0]

+3 ± 6 [4]

−2 ± 11 [0]

 

RV sensing

 

−1 ± 6 [−1]

−2 ± 12 [−2]

−4 ± 14 [−1]

 

RV PTH

 

+2 ± 8 [0]

0 ±12 [0]

7 ± 24 [0]

 

RV PImp

 

−2 ± 2 [−2]

0 ± 6 [0]

−2 ± 7 [0]

 

Battery status

 

0 ± 0 [0]

0 ± 0 [0]

0 ± 1 [0]

 
  1. Grey columns: relevant FUs for endpoint calculation.
  2. *Oneway ANOVA.
  3. +p < 0.05 when comparing with pre MRI values (paired t-test).