Skip to main content

Table 3 Diagnostic performance of each imaging component in detecting MI: Sensitivity, Specificity, Positive Predictive Value, Negative Predictive Value

From: MultiContrast Delayed Enhancement (MCODE) improves detection of subendocardial myocardial infarction by late gadolinium enhancement cardiovascular magnetic resonance: a clinical validation study

 

LGE1

LGE2

MCODE

LGE3

Sensitivity*

83%

85%

100%

85%

Specificity**

92%

92%

88%

92%

Positive predictive value (PPV)**

95%

95%

94%

95%

Negative predictive value (NPV)*

75%

77%

100%

77%

Accuracy

86%

87%

96%

87%

  1. *Sensitivity of LGE1 compared to MCODE, p = 0.008; NPV comparison, p = 0.005.
  2. *Sensitivity of LGE2/LGE3 compared to MCODE, p = 0.02; NPV comparison, p = 0.009.
  3. **Specificity and PPV of LGE1/LGE2/LGE3 compared to MCODE, p = not statistically significant.