Skip to main content

Table 2 In-vitro results

From: High resolution slice-selective Fourier Velocity Encoding using spiral SENSE with velocity unwrap

 

Echo

lh-PCMR

hr-PCMR

FVE

Peak velocity (cm/s)

441±144

375±133^

398±136^

443±144

Bias* (cm/s)

-

-66

-42

+3

Limits of Agreement* (cm/s)

-

-26 to -105

-10 to -75

+17 to -12

Correlation coefficient* (r)

-

0.9926

0.9949

0.9987

  1. * Calculated with echo ^ Value is significantly different (ANOVA) from echo (P < 0.05)