In-vivo distortion of through-plane flow by spiral phase-contrast imaging
Journal of Cardiovascular Magnetic Resonance volume 14, Article number: W61 (2012)
Summary
To explain previously unrecognised consequences of two sources of phase curvature over the vessel cross-section in spiral imaging i.e. off-resonance and the velocity-encoded phase-shift.
Background
The effects of off-resonance frequency errors during spiral readouts [Yudilevich and Stark, 1987] are known. Here we explain previously unrecognised intra-voxel dephasing consequences of two sources of phase curvature over the vessel cross-section in through-plane flow imaging using spirals i.e. off-resonance and the velocity-encoded phase-shift, including the consequences for in-vivo measurements.
Methods
Laminar through-plane flow phantom (50cm/s) and popliteal artery studies were acquired at Venc=50cm/s with initial off-resonance offsets ± 0 10, 20, 40Hz representing <1ppm at 1.5T. Reference (“Ref”) and velocity-encoded (“Vel”) magnitude and phase images were obtained (as a cine in-vivo). Spiral FOV was 150 mm, 1mm resolution, duration 25.7ms; TE/TR 4.0/32.7 ms, FA 30°, 4 interleaves.
Results
Figure 1 shows magnitude and phase images for Ref and Vel scans with corresponding velocity maps (VM). For low flow rates the off-resonance blurring at +40Hz and -40Hz is similar for both Ref. and Vel. images. However, high velocities cause an apparent “implosion” (Figure 1-I) and “explosion” (Figure 1-E) of the vessel for opposite off-resonance frequencies. From the distorted VMs, peak velocity was measured at 37.8, 48.6, 56.3cm/s for -40, 0, 40Hz off-resonance respectively.
For theoretical explanation, Figure 2a) depicts the ideal diametrical phase line profile across a tube velocity-compensated spiral imaging, red line = on-resonance and blue/green lines = phase curvature induced by ±40Hz off-resonance error. Figure 2b)-red shows velocity-encoded phase over the vessel with parabolic flow. Figure 2b)-blue shows the consequence of their addition: the increased radial slope worsens intra-voxel dephasing in all but the central pixels of a laminar flow (“implosion” effect, Figure 1-I). In Figure 2b)-green the opposing phase curvatures lead to a cancellation of the radial phase slope at some radius, which can lie outside the true lumen (“explosion” effect, Figure 1-E).
Conclusions
The distortion of the velocity distribution over the in-vivo vessel distorts peak velocity by ~20% at -40Hz off-resonance at these sequence parameters. Separate tests eliminated through-plane gradient fields as a cause, including eddy-current effects after the velocity-encoding pulses.
These effects on magnitude images and velocity distributions at <1ppm off-resonance are potentially difficult for >20ms spiral readouts in small vessel applications at least, perhaps more so near B0-distortions such as lungs. Shorter spirals and avoiding large intra-voxel radial phase shear are some-what incompatible with rapid flow work.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Pierce, I.T., Gatehouse, P.D. & Firmin, D.N. In-vivo distortion of through-plane flow by spiral phase-contrast imaging. J Cardiovasc Magn Reson 14 (Suppl 1), W61 (2012). https://doi.org/10.1186/1532-429X-14-S1-W61
Published:
DOI: https://doi.org/10.1186/1532-429X-14-S1-W61