Skip to main content

Table 1 SHILO imaging parameters

From: SHILO, a novel dual imaging approach for simultaneous HI-/LOw temporal (Low-/Hi-spatial) resolution imaging for vascular dynamic contrast enhanced cardiovascular magnetic resonance: numerical simulations and feasibility in the carotid arteries

Imaging parameters AIF VW
Field of view (FOV) 160 mm × 160 mm
In-plane spatial resolution 0.5 mm × 2 mm 0.5 mm × 0.5 mm
N of slices 1
Slice Thickness 3 mm
Acquisition matrix 320 × 80 320 × 320
N of acquisition segments 1 4
Turbo field echo (TFE) factor 80
Repetition time (TR) 10 ms
Echo time (TE) 4 ms
Saturation delay None 800 ms
Flip angle 15 25
K-space linear centric
Time resolution 1.6 s 6.4 s