Skip to main content

Table 3 Results of left and right ventricular (LV, RV) functional analysis performed by the two CMR sequences, self-gated FLASH and self-gated UTE as well as by echocardiography

From: Cardiac-respiratory self-gated cine ultra-short echo time (UTE) cardiovascular magnetic resonance for assessment of functional cardiac parameters at high magnetic fields

 

Echocardio-graphy§

CMR Self-gated cine FLASH§

CMR Self-gated cine UTE§

p-values FLASH-UTE/FLASH-Ultrasound§

LV morphology and function

Mass [mg]

91 ± 11

83.7 ± 5.8

83.6 ± 3.0

0.98/0.20

Stroke volume (SV) [μl]

35.0 ± 5.4

32.3 ± 6.9

32.3 ± 6.5

0.99/0.47

Ejection fraction (EF) [%]

54.7 ± 4.8

72.2 ± 3.2

71.8 ± 2.2

0.78/0.000022

Cardiac output (CO) [ml/min]

16.6 ± 3.2

15.6 ± 3.8

15.6 ± 3.8

0.99/0.63

RV morphology and function

Mass [mg]

-

15.0 ± 4

19.3 ± 3.7

0.10

Stroke volume (SV) [μl]

31.7 ±7.5 (Doppler)

22.9 ± 3.5

22.6 ± 3.9

0.89/0.026

Ejection fraction (EF) [%]

-

72.4 ± 6.1

69.6 ± 3.2

0.34

Cardiac output (CO) [ml/min]

13.2 ± 3.2 (Doppler)

11.0 ± 2.2

10.9 ± 2.1

0.89/0.20

  1. §Values are given as mean ± standard deviation values averaged over six healthy C57BL/6 mice.