Figure 2 | Journal of Cardiovascular Magnetic Resonance

Figure 2

From: Modified look-locker inversion recovery T1 mapping indices: assessment of accuracy and reproducibility between magnetic resonance scanners

Figure 2

Accuracy measurements for different vials. Bar graph representation comparing MOLLI T1 accuracy measurements across four different vials. Accuracy is reported as the percentage difference between MOLLI and IR-SE. Data is presented as least square means ± standard error. Smaller numbers represent better accuracy. The two post-contrast vials with lower T1 times had better accuracy than the two pre-contrast vials with longer T1 values. The accuracy of partition coefficient is in the middle of pre and post-contrast vials.