Table 4 MOLLI T1 and partition coefficient accuracy and reproducibility across field strengths and protocols

From: Modified look-locker inversion recovery T1 mapping indices: assessment of accuracy and reproducibility between magnetic resonance scanners

  Field strength Protocol
  1.5T 3T p-value 17HB 11HB p-value
Accuracy (% error)
T1 accuracy 6.3 ± 0.3 10.8 ± 0.2 <0.0001 8.3 ± 0.3 8.8 ± 0.3 0.18
Partition coefficient accuracy 8.8 ± 0.6 8.0 ± 0.5 0.11 8.3 ± 0.6 8.5 ± 0.6 0.57
Reproducibility (% mean difference)
T1 reproducibility 5.3 ± 0.3 3.9 ± 0.2 <0.0001 4.5 ± 0.2 4.7 ± 0.2 0.43
Partition coefficient reproducibility 3.8 ± 0.3 2.8 ± 0.2 0.02 2.7 ± 0.3 4.0 ± 0.3 0.003
  1. Note: The accuracy and reproducibility of MOLLI on field strength and protocol are reported. Accuracy is reported as the percentage difference between MOLLI and IR-SE. Scan/rescan reproducibility is reported as the relative percent mean difference between two MOLLI scans. Data is presented as least square means ± standard error. Smaller numbers represent better accuracy or reproducibility.