Table 1 Acquisition parameters for real-time SSFP CMR

From: Real-time cardiovascular magnetic resonance at 1.5 T using balanced SSFP and 40 ms resolution

 

CMR I

CMR II

Imaging time/ms

41

40

Frame rate/s-1

25

25

Maximum gradient/mT m-1

40

30

Slew rate/mT m-1 ms-1

200

125

Resolution/mm3

1.8 × 1.8 × 6.0

1.8 × 1.8 × 8.0

Field-of-view/mm2

256 × 256

256 × 256

Base resolution

144

144

Acquired spokes

15

13

Nyquist undersampling factor

15

17.4

Repetition time/ms

2.72

3.08

Echo time/ms

1.36

1.54

Flip angle/degree

35

35

Bandwidth/Hz Pixel-1

1930