Skip to main content

Table 3 Intra- and interstudy limits-of-agreement

From: Novel insight into the detailed myocardial motion and deformation of the rodent heart using high-resolution phase contrast cardiovascular magnetic resonance

 

Intrastudy variability

Interstudy variability

Radial velocity

0.01 ± 0.32 cm/s

0.07 ± 2.30%venc

−0.06 ± 0.75 cm/s

−0.43 ± 5.40%venc

Circ. velocity

0.10 ± 0.35 cm/s

0.72 ± 2.52%venc

0.06 ± 0.94 cm/s

0.43 ± 6.76%venc

Long. velocity

0.13 ± 0.51 cm/s

0.94 ± 3.67%venc

0.12 ± 1.15 cm/s

0.86 ± 8.27%venc

Sc

−0.16 ± 0.89%

−0.71 ± 1.67%

Sc (single motion tracking)*

−0.15 ± 0.96%

−0.78 ± 2.31%

  1. Velocity variability shown in absolute values (cm/s) and percentage of the applied venc. Strain variability is reported in absolute values (%strain). Sc = Circumferential strain. N = 43x6 = 258. Values are mean ± 1.96*SD.
  2. *: Sc was also calculated from motion paths using only end-diastole as temporal origin, as described in [23], without our proposed extension of also including end-systole as temporal origin.