Table 1 Reproducibility according to type of MRI scanner

From: Quantification of myocardial scar assessed by late gadolinium enhancement CMR in the multi-ethnics study of atherosclerosis: comparisons of 7 different methods

 

Scan Type

  
 

Siemens (n = 109)

GE (n = 23)

  
 

Bias

95% loa

Bias

95% loa

  

Method of Quantification by Scar Pattern

Ischemic

BCT

-0.27

-11.8, 11.2

0.96

-5.7, 7.6

  

FWHM

0.98

-8.5, 10.4

0.31

-8.1, 8.8

  

8SD

-0.09

-12.4, 12.2

2.8

-5.1, 10.7

 

Non-Ischemic

BCT

-0.24

-3.2, 2.7

0.17

-2.5, 2.8

  

8 SD

-0.67

-9.3, 8.3

1.1

-1.9, 4.1

  1. Comparison of Bland Altman analysis for average difference (bias) and the limits of agreement between the semi-automated method and three automatic techniques according to the scanner type where they were performed. Only the three automatic techniques that did not show significant difference from the semi-automated in the overall analysis were included in this subanalysis. BCT= Background corrected technique, 8 SD= 8 standard deviations, FWHM= Full Width at Half Maximum