Skip to main content
  • Oral presentation
  • Open access
  • Published:

Improved fat signal suppression for coronary MRA at 3T using a water-selective adiabatic T2-Prep technique

Background

In MRI, fat signal suppression is mandatory for unambiguous visualization of cardiac anatomy but remains challenging at high field. Fat saturation usually directly precedes the imaging portion of a cardiac MRI sequence. It is often used in conjunction with other magnetization preparation modules, such as T2 Preparation (or T2-Prep), to improve blood/myocardium contrast. We propose a water-selective, adiabatic T2-Prep that combines T2-weighting with an intrinsic fat signal attenuation. Preliminary in vivo results are discussed.

Methods

The first radiofrequency (RF) pulse of a +90°,180°,180°,-90° adiabatic T2-Prep was reduced in bandwidth, from 1250 Hz to 285 Hz. Water is thus excited but off-resonant fat magnetization is left longitudinal. Meanwhile, the bandwidth of the final RF pulse (-90°) remains large (1250 Hz), encompassing both water and fat frequencies. It thus restores the magnetization of T2-prepared water, while fat magnetization is tipped down and then spoiled. The RF excitation angles of the first and last pulse were experimentally increased to ±120°, to further reduce fat signal via inversion recovery. To demonstrate how these modifications can easily combine with other fat saturation strategies, images were acquired with and without conventional frequency selective fat saturation (CHESS), for both the water-selective adiabatic T2-Prep and the unmodified adiabatic T2-Prep. Volume targeted 3D imaging of the right coronary artery was performed for all 4 sequences in 6 healthy adults. All images were acquired on a 3T Siemens TRIO using a navigator- and cardiac-gated segmented k-space Cartesian gradient echo sequence, with FoV 360x258 mm, matrix size 240x216, 1.5 mm slice thickness, 15 k-space lines/heartbeat, TE T2-Prep = 40 ms, RF excitation angle 15°, TE/TR/TAcq=2.37/5.37/80.55 ms. Images were reformatted and analyzed using semi-automated vessel tracking software (Soap-Bubble). Fat suppression efficacy was compared using vessel sharpness measurements and signal-to-noise ratio (SNR) quantification in selected regions (abdominal fat, epicardial fat, blood, myocardium).

Results

Sample images are shown in Figure 1, with corresponding SNR measurements in Table 1. When no complementary fat saturation was used, the water-selective adiabatic T2-Prep reduced abdominal and epicardial fat signals by 36% and 16% (p<0.001 and p<0.05) and improved vessel sharpness from 43.8% to 47.3% (p<0.05), as compared to the unmodified T2-Prep. When a CHESS pulse was added prior to imaging, the water-selective adiabatic T2-Prep reduced abdominal and epicardial fat signals by a further 54% and 30% (p<0.002 and p<0.02) and improved vessel sharpness from 68.4% to 72.7% (p<0.05), as compared to unmodified T2-Prep + CHESS. Blood and myocardium SNRs were not significantly affected.

Figure 1
figure 1

Increased fat signal suppression of the variable bandwidth (BW) water selective adiabatic (WSA) T2-Prep technique as compared to conventional broadband adiabatic T2-Prep. Fat suppression is improved both with and without an additional frequency-selective fat saturation (CHESS) pulse.

Table 1 Mean SNR (± one S.D.) for conventional broadband adiabatic T2-Prep vs. water selective adiabatic T2-Prep, both with or without an additional frequency-selective fat saturation (CHESS) pulse.

Conclusions

A water-selective adiabatic T2 Preparation module significantly improves fat saturation in 3T coronary MRA and should be considered as a potential addition to conventional fat saturation strategies.

Funding

N/A

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Coristine, A.J., van Heeswijk, R.B. & Stuber, M. Improved fat signal suppression for coronary MRA at 3T using a water-selective adiabatic T2-Prep technique. J Cardiovasc Magn Reson 15 (Suppl 1), O5 (2013). https://doi.org/10.1186/1532-429X-15-S1-O5

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-15-S1-O5

Keywords