Figure 1From: Frequency drift during intensive SSFP scanning: implications and solution for 3T neonatal CMRFrequency drift test in phantoms during prolonged SSFP: reformatted views (phase-slice direction) of a multi-2D stack acquisition in (a) fast reference scan - no drift, (b) neonatal optimised protocol and (c) added frequency stabilisation. Single slice view from (d) a 3D cine SSFP protocol also show a progressive frequency drift through repeated scan acquisitions 1-3.Back to article page