From: Semi-automatic inline calculation of left ventricular function using cardiac MRI (CMR)
Parameters | Automatic | Semi-automatic | Manual |
---|---|---|---|
EF (%) | 50.60 ± 10.46 | 60.20 ± 9.83 | 61.80 ± 9.53 |
EDV (mL) | 157.30 ± 55.89 | 140.60 ± 49.70 | 143.30 ± 51.01 |
ESV (mL) | 79.60 ± 40.74 | 58.30 ± 33.10 | 57.0 ± 32.54 |
SV (mL) | 77.80 ± 28.22 | 82.20 ± 25.44 | 86.30 ± 27.64 |
CO (L/min) | 4.90 ± 1.68 | 5.20 ± 1.57 | 5.50 ± 1.69 |
Mean Measuring Time (min:sec) | 0 | 2:20 | 5:11 |