Figure 12

Phase sensitive inversion recovery (PSIR) fitting uses a 3-parameter model, whereas magnitude IR fitting using a multi-fitting approach estimates 3-parameters plus the zero-crossing. The multi-fitting magnitude IR fitting approach is prone to errors in estimating the zero-crossing in situations where the zero-crossing is close to the measured inversion times leading to a significant loss of precision for specific values of T1 and RR for a given protocol.