Figure 1 | Journal of Cardiovascular Magnetic Resonance

Figure 1

From: Safe performance of magnetic resonance of the heart in patients with magnetic resonance conditional pacemaker systems: the safety issue of the ESTIMATE study

Figure 1

Device related imaging artifacts. A: Vertical long axis view showing subtle susceptibility artefacts (short arrow) around the ferromagnetic right ventricular lead (long arrow). In addition, there are prominent off-resonance stripe artifacts in the left ventricle (small arrows) with no impairment of endocardial border detection. B: Short axis view with the typical signal void caused by the RV lead (short arrow) in the inferior wall not interfering with the endocardial border. Note the large signal void and circular stripe artifacts caused by the generator (long arrow) with no interference with the heart.