From: MRI-compatible voltage-based electroanatomic mapping system for 3T MR-guided cardiac electrophysiology: swine validations

Figure 1

EAM maps acquired (A) outside MRI versus (B) inside the 3T MRI. Three simultaneously navigated catheters (LA, RA#1 and RA#2) are shown. Upper images show EAM maps. Lower images show surface ECG, intra-cardiac EGM (RA#1, RA#2 and LA catheter tip electrodes), as well as intra-aortic IBP traces. QRS denotes start of ECG cycle. V and A denote Ventricular and Arterial signals, respectively. A strong superimposed ventricular signal was observed on all atrial intra-cardiac ECG traces (extreme-right red vertical line in (B)) acquired inside the MRI, a result of the strong 3T Magnetohydrodynamic voltage.