Skip to main content

Table 1 Comparison of results of SE, IR-rttfl and MOLLI in phantom

From: T1 mapping of myocardium using inversion recovery cine at 3.0T

 

Spin Echo

IR- rttfl

MOLLI

Phantom number

T2(ms)

T1(ms)

T1(ms)

T1(ms)

1

346.7 ± 2.6

1505 ± 4

1518 ± 50

1403 ± 8

2

320.5 ± 0.9

1247 ± 3

1240 ± 35

1197 ± 8

3

284.0 ± 0.8

867 ± 3

870 ± 38

856 ± 7

4

260.6 ± 0.5

696 ± 2

693 ± 38

692 ± 7

5

177.0 ± 0.5

316 ± 1

313 ± 31

316 ± 12

6

160.8 ± 0.6

271 ± 1

252 ± 38

279 ± 7