Duration | 4.5-7 minutes |
---|---|
Hardware | Phillips Achieva 1.5T magnet; 5-channel phased array coil; Max. parallel acceleration: 2 (2D); 2X1.5 (3D) |
Acquisition Parameters | Axial or sagittal acquisition; respiratory triggered, retrospective EKG gated, TR/TE/α = 3/1.5/60; voxel size 1.5-2.5; SENSE 2.6-3.6; mean temporal resolution = 35 ms |
Imaging Targets & Grading Methodology | Quantitative LV volumetry (3D vs 2D cine SSFP): LVEDV, LVESV, LVEF Vessel Measurements (3D cine SSFP vs 2D cine SSFP and Gd-3D MRA): Maximal intersinus aortic root, ascending aorta, aortic isthmus, MPA and LPA diameter Cardiac Morphology Qualitative scoring (3D vs 2D cine SSFP): Atrial morphology/ function (0-8): Edge definition (0-2), Blood myocardial contrast (0-2), Atrial septum (0-2), Crista terminalis (0-2), Aortic valve morphology/ function (0-8): Annulus (0-2), Number of leaflets (0-2), Valve motion (0-2), Valve coaptation (0-2) Tricuspid valve morphology/ function (0-10): Annulus (0-2), Number of leaflets (0-2), Valve motion (0-2), Valve coaptation (0-2), Suspensory apparatus (0-2) Left Ventricular morphology (0-15): Edge definition (0-5), Blood myocardial contrast (0-5), Interslice alignment (0-5) Extra-cardiac Vasculature Qualitative Scoring (3D cine SSFP vs Gd-3D MRA): Aorta (0-8): Edge definition (0-2), Blood pool homogeneity (0-2), First order branches (0-2), Second order branches (0-2) Pulmonary arteries (0-10): Edge definition (0-2), Blood pool homogeneity (0-2), First order branches (0-2), Second order branches (0-2), Arteriovenous separation (0-2) Pulmonary veins (0-10): Edge definition (0-2), Blood pool homogeneity (0-2), First order branches (0-2), Second order branches (0-2), Arteriovenous separation (0-2) SVC/ Azygous vein (0-8): Edge definition (0-2), Blood pool homogeneity (0-2), First order branches (0-2), Second order branches (0-2) |