Skip to main content
Figure 2 | Journal of Cardiovascular Magnetic Resonance

Figure 2

From: Joint myocardial T1 and T2 mapping

Figure 2

T1 and T maps from a healthy subject, acquired using the proposed technique, as well as SASHA T1 mapping, and conventional T2 mapping using 4 T2prep echo times. Both the T1 and T2 maps generated jointly with the proposed method are similar to the individual maps with similar magnetization preparations. The myocardial T1 and T2 values in the septum were 1211 ± 82 ms (SASHA T1), 1210 ± 92 ms (Joint T1-T2), 49.0 ± 5.8 ms (conventional T2) and 47.3 ± 6.5 ms (Joint T1-T2) for each technique. The methods generated with the proposed method were acquired in the same time as each individual map, and are jointly registered by design.

Back to article page