Positive contrast spiral imaging of a nitinol guidewire
Journal of Cardiovascular Magnetic Resonance volume 17, Article number: Q15 (2015)
Background
The clinical translation of MRI-guided cardiovascular catheterization has been limited by the unavailability of devices that are both visible and safe under MRI. In particular, rigid metallic guidewires are essential for most catheterization procedures and are at risk of heating during MR imaging [1]. Here we present an MRI method that simultaneously improves the visualization of commercially available nitinol guidewires and minimizes RF induced heating.
Methods
RF-efficient gradient echo spiral imaging was chosen to minimize heating (8 interleaves, TE/TR = 0.86/10ms, flip angle = 10°). Through-slice dephasing generated a positive contrast "device image" [2], exploiting local field inhomogeneity such that the metallic guidewire appears hyperintense with background signal suppressed. An anatomical image and a device image were interleaved in alternating frames. Image processing (signal thresholding and selection of elongated structures) was performed on the device image to isolate the guidewire signal from other sources of field inhomogeneity.
Imaging was performed on a 1.5T MRI scanner (Aera, Siemens, Erlangen, Germany). MRI-guided left heart catheterization was performed in a pig using a 0.035" commercially available nitinol guidewire (Nitrex, Covidien, Plymouth, MN). The RF induced temperature rise at the tip of an insulated nitinol rod during MR imaging was measured in the ASTM 2182 gel phantom using a fiberoptic temperature probe (OpSense, Quebec, Canada).
Results
A pair of anatomical and device images were generated with a temporal resolution of 160ms (80ms per image), or 6.25 frames/s. Image processing occurred in real-time and a color overlay of the device on the anatomy was displayed to the operator for guidance during catheterization (Figure 1). Isolation of the guidewire signal from the background was most challenging around the aortic arch.
During 1 minute of continuous scanning in vitro, our spiral imaging method generated 0.47°C at the tip of a nitinol rod, compared to 37.2°C of heating generated using standard Cartesian bSSFP imaging (Figure 2).
Conclusions
This method simultaneously improves guidewire visualization using a real-time color overlay and minimizes heating using RF-efficient imaging. Thus, this method may permit the safe and effective use of standard commercially available metallic guidewires for MRI-guided cardiovascular catheterizations.
Funding
This work was supported by the NHLBI DIR (Z01-HL006039-01, Z01-HL005062-08).
References
Konings MK: Heating around intravascular guidewires by resonating RF waves. JMRI. 2000, 12 (1): 79-85. 10.1002/1522-2586(200007)12:1<79::AID-JMRI9>3.0.CO;2-T.
Seppenwoolde JH: Passive Tracking Exploiting Local Signal Conservation: The White Marker Phenomenon. MRM. 2003, 50: 784-790. 10.1002/mrm.10574.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Campbell-Washburn, A.E., Rogers, T., Basar, B. et al. Positive contrast spiral imaging of a nitinol guidewire. J Cardiovasc Magn Reson 17 (Suppl 1), Q15 (2015). https://doi.org/10.1186/1532-429X-17-S1-Q15
Published:
DOI: https://doi.org/10.1186/1532-429X-17-S1-Q15