From: Free-breathing myocardial T1 mapping using magnetization-prepared slice interleaved spoiled gradient echo imaging

Figure 1

Accuracy, precision and reproducibility of the proposed sequence obtained in Monte Carlo simulation (20,000 repetitions, fixed T2 of 50 ms, SNR corresponding to 50 in the ∞ image) (a) and phantom experiments (set of vials with NiCl2 doped agarose, 15 repetitions of the sequence) (b). Results were compared to the MOLLI (5-(3)-3 scheme) sequence. Accuracy was measured in each vial as the difference between spin echo T1 measurements and the average T1 over all 15 repetitions. Precision was measured in each vial as the average (over all 15 repetitions) of the standard deviation of T1 within a vial. Reproducibility was measured in each vial as the standard deviation (over all 15 repetitions) of the mean T1 within a vial. Improved accuracy and similar precision were achieved using the proposed sequence in both simulations and phantom experiments. T1 mapping reproducibility was slightly decreased with the proposed sequence.