Skip to main content

Table 1 Diagnostic Performance in Canines

From: Flow-Independent Dark-blood DeLayed Enhancement (FIDDLE): validation of a novel black blood technique for the diagnosis of myocardial infarction

 

Sensitivity

Specificity

Accuracy

Overall

FIDDLE

97% (95/98)

92% (35/38)

96% (130/136)

DE-CMR

81% (79/98)

95% (36/38)

85% (115/136)

p-value

< 0.001

0.65

0.001

Subendocardial MI (transmurality < 25%)

FIDDLE

98% (44/45)

92% (35/38)

95% (79/83)

DE-CMR

71% (32/45)

95% (36/38)

82% (68/83)

p-value

< 0.001

0.65

0.008