Table 1 Comparison between GRAPPA and SENSE accelerated cardiac ASL

From: Myocardial ASL-CMR perfusion imaging with improved sensitivity using GRAPPA

   Global Per-segment
  FOV (cm) MBF (ml/g/min) PN (ml/g/min) TSNR N MBF (ml/g/min) PN (ml/g/min) TSNR N
GRAPPA 20.1 ± 2.8 1.73 ± 0.22 0.21 ± 0.10 11.48 ± 8.85 532.0 ± 132.7 1.97 ± 0.85 0.44 ± 0.22 5.44 ± 3.04 133.6 ± 35.7
SENSE 26.6 ± 3.1 1.75 ± 0.85 0.60 ± 0.34 3.41 ± 1.38 269.6 ± 43.2 1.78 ± 1.04 0.73 ± 0.55 3.26 ± 3.15 67.6 ± 13.5
p-value 0.0003 0.9681 0.0289 0.0086 0.0061 0.3230 0.0011 0.0013 2.6 × 10-12
  1. Global analysis was performed on the entire left ventricular myocardium. Per-segment analysis divided the myocardium into 6 regional segments. Values are reported as mean ± standard deviation across all subjects. (FOV) field of view, (MBF) myocardial blood flow, (PN) physiological noise, (TSNR) temporal signal-to-noise, (N) number of pixels within each segment.