Motion Corrected 3D Whole-Heart Vessel Wall Imaging
Journal of Cardiovascular Magnetic Resonance volume 18, Article number: P323 (2016)
Background
Coronary atherosclerosis is not necessarily stenotic, due to outward remodeling of the vessel wall. Plaque burden correlates with risk of coronary disease (Kubo et al, J Am Coll Cardiol 2007) and direct visualization is desired. A 3D flow independent approach for vessel wall imaging was proposed recently (Andia et al, MRM 2013), based on subtraction of data with (T2prep(+)) and without (T2prep(-)) a T2-preparation prepulse. However, T2prep(+) and T2prep(-) data are only accepted when both are within the acquisition window given by the diaphragmatic navigator, leading to increased scan times. Here, we propose to accelerate the acquisition by performing respiratory motion correction on T2prep(-) and T2prep(+) with 100% scan efficiency. This is achieved with a beat-to-beat translational correction combined with a bin-to-bin non-rigid correction.
Methods
Data are acquired using an interleaved scanning framework (Henningsson et al, MRM 2014). T2prep(-) and T2prep(+) are acquired interleaved every other heartbeat and preceded by a 2D coronal low resolution image navigator (iNAV) (Figure 1a). Translational foot-head (FH) and right-left (RL) motion is estimated from the iNAV and used to bin data into FH respiratory motion states. Within each bin, 2D translational correction is applied directly in k-space (Figure 1b). Bins are reconstructed with parallel imaging and non-rigid motion is subsequently estimated via image registration. Non-rigid motion between bins is corrected using the General Matrix Description (GMD) (Batchelor et al, MRM 2005).
Eight healthy volunteers were scanned under free-breathing on a 1.5T Philips scanner using a 32-channel coil. Image data was acquired with an ECG-triggered 3D balanced SSFP: coronal slices, flip angle = 70°, TR/TE = 4.6/2.3 ms, FOV = 300 × 300 × 100 mm3, resolution = 1 × 1 × 2 mm3. iNAV data was acquired with a 2D Golden radial spoiled gradient echo: coronal slice, flip angle = 5°, TR/TE = 2.4/1.1 ms, FOV = 300 × 300 mm2, resolution = 4 × 4 mm2. Data was reconstructed with no motion correction (NMC), 2D beat-to-beat translation correction (TC) only and with the proposed approach (TC+GMD).
Results
Reformatted T2prep(+), vessel wall images and intensity profiles along the left and right coronary arteries (LCA and RCA) are shown in Figure 2. Coronary vessels are not visible in NMC due to respiratory motion. A significant improvement is obtained in TC: RCA vessel wall becomes visible, but some residual blurring remains in the LCA. TC+GMD removes further motion artifacts leading to clearer delineation of the vessel wall in both coronary arteries. All volunteers showed visual vessel wall improvement after TC+GMD.
Conclusions
Good quality vessel wall imaging is achieved with the proposed approach with 100% scan efficiency. The proposed motion correction shows visible improvements over translational correction only, leading to a better delineation of the coronary vessel wall. Future work will focus on additional acceleration using compressed sensing.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Lima da Cruz, G.J., Atkinson, D., Henningsson, M. et al. Motion Corrected 3D Whole-Heart Vessel Wall Imaging. J Cardiovasc Magn Reson 18 (Suppl 1), P323 (2016). https://doi.org/10.1186/1532-429X-18-S1-P323
Published:
DOI: https://doi.org/10.1186/1532-429X-18-S1-P323