Skip to main content

Table 1 Summary of change in phase offsets for sagittal slice orientation at flow-encoding velocity of 150 cm/s (TE =1.85 ms)

From: Analysis of temperature dependence of background phase errors in phase-contrast cardiovascular magnetic resonance

 

1storder phase offset at 10 cm from iso-center [% venc]

0th order phase offset [% venc]

 

X → X

Y → Y

Z → Z

X → B0

Y → B0

Z → B0

T =0 min

−2.1 ± 0.2

18.8 ± 0.2

3.1 ± 0.2

−6.8 ± 0.3

0.9 ± 0.3

−1.4 ± 0.3

T =12 min

−8.8 ± 0.2

22.2 ± 0.2

9.1 ± 0.2

−4.3 ± 0.3

0.9 ± 0.3

−0.3 ± 0.3

Difference

−6.7 ± 0.3

3.4 ± 0.3

6 ± 0.3

2.5 ± 0.4

0 ± 0.4

−1.1 ± 0.4