Figure 8 | Journal of Cardiovascular Magnetic Resonance

Figure 8

From: Simultaneous three-dimensional myocardial T1 and T2 mapping in one breath hold with 3D-QALAS

Figure 8

Effect of deliberately applied arrhythmias on phantom T1 measurements. Gaussian noise distributions with three different widths (5%, 10% and 15%) were used to change the length of the cardiac cycle, initially 60 beats per minute (bpm), in order to simulate atrial fibrillation. Deviation in percentage from measurements with 60 bpm and 0% noise can be seen as a function of T1 values measured with 3D-QALAS at a heart rate of 60 bpm with 0% noise.