Skip to main content

Table 2 Inter-observer reproducibility (initial scanning session) and inter-study reproducibility (observers 1 and observer 2) of cross-sectional area, RABF, mean velocity, resistive index (RI) and pulsatility index (PI)

From: Inter-study reproducibility of interleaved spiral phase velocity mapping of renal artery haemodynamics

Reproducibility

Area (mm 2 )

RABF per kidney (ml/min)

Mean velocity (mm/s)

RI

PI

Inter-observer

−7.2+/−2.1**

38.5+/−20.0**

−32.3+/−8.6**

0.01 + −0.01*

0.07+/−0.08*

Inter-study (observer 1)

−2.0+/−2.4

17.9+/−44.8

−6.7+/−31.1

0.00+/−0.04

−0.03+/−0.17

Inter-study (observer 2)

−2.6+/−4.1

24.2+/−59.0

−5.5+/−36.3

0.00+/−0.05

−0.01+/−0.21

  1. (**p < .001, *p < .01).
  2. All data are presented as mean (+/− standard deviation) of the signed differences between measurements.