Skip to main content

Table 2 Comparison between the reference LA volumes the CS-cineCMR combined with the 3D non-model-based and the model-based bi-plane area-length reconstruction. Acquisition strategies A1 to E3 are explained in Fig. 1

From: Single breath-hold 3D measurement of left atrial volume using compressed sensing cardiovascular magnetic resonance and a non-model-based reconstruction approach

 

Acquisition

CS-cineCMR 3D non-model based Recon.

Reference

Diff

Diff [%]

 

Volume [ml]

Volume [ml]

Volume [ml]

Oval big

A1

51.0

54.0

−3.0

−5.8 %

A2

50.9

−3.1

−6.0 %

A3

52.8

−1.2

−2.3 %

Oval small

B1

45.7

51.0

−5.3

−11.0 %

B2

46.4

−4.6

−9.5 %

B3

46.5

−4.5

−9.2 %

Oval oblique

C1

55.3

61.0

−5.7

−9.8 %

C2

55.3

−5.7

−9.9 %

C3

54.4

−6.6

−11.5 %

Spherical

D1

48.2

55.0

−6.8

−13.3 %

D2

47.2

−7.8

−15.3 %

D3

49.8

−5.2

−10.0 %

Curved

E1

65.8

70.0

−4.2

−6.1 %

E2

67.5

−2.5

−3.6 %

E3

65.3

−4.7

−6.9 %

All

-

53.5 ± 7.3

58.2 ± 7.0

−4.7 ± 1.8

−8.7 ± 3.5 %

Oval big

A1

52.67

54.0

−1.33

−2.5 %

A2

50.29

−3.71

−7.1 %

A3

46.51

−7.49

−14.9 %

Oval small

B1

48.38

51.0

−2.62

−5.3 %

B2

47.76

−3.24

−6.6 %

B3

45.03

−5.97

−12.4 %

Oval oblique

C1

55.92

61.0

−5.08

−8.7 %

C2

57.28

−3.72

−6.3 %

C3

42.57

−18.43

−35.6 %

Spherical

D1

50.10

55.0

−4.90

−9.3 %

D2

51.66

−3.34

−6.3 %

D3

49.96

−5.04

−9.6 %

Curved

E1

61.02

70.0

−8.98

−13.7 %

E2

63.23

−6.77

−10.2 %

E3

54.29

−15.71

−25.3 %

All

-

51.8 ± 5.8

58.2 ± 7.0

−6.4 ± 4.8

−11.5 ± 8.5 %