Skip to main content

Table 4 Effect of CMR scanner and acquisition protocol

From: Estimation of aortic pulse wave transit time in cardiovascular magnetic resonance using complex wavelet cross-spectrum analysis

archPWVWU

 

Avg by 2

Avg by 3

Avg by 4

Linear fit slope

Group 1

1.02

0.94

0.91

Group 2

1.05

0.96

0.88

Bland-Altman mean bias [limits] (m/s)

Group 1

0.32 [-1.30 to 1.95]

0.30 [-1.54 to 2.16]

0.37 [-1.68 to 2.42]

Group 2

0.07 [-1.58 to 1.73]

-0.06 [-2.51 to 2.38]

-0.16 [-3.17 to 2.84]

Correlation r(p)

Group 1

0.92 (<0.001)

0.90 (<0.001)

0.87 (<0.001)

Group 2

0.96 (<0.001)

0.90 (<0.001)

0.85 (<0.001)

  1. For the wavelet-based method, aortic pulse wave velocity (archPWVWU) estimated from the original ascending and descending aortic flow curves, was used as a reference for comparisons against archPWVWU estimated from curves with lower temporal resolution obtained by averaging time points by blocks of 2 (Avg by 2), 3 (Avg by 3) and 4 (Avg by 4) in both Group 1 and Group 2