Skip to main content
Fig. 7 | Journal of Cardiovascular Magnetic Resonance

Fig. 7

From: Gadolinium free cardiovascular magnetic resonance with 2-point Cine balanced steady state free precession

Fig. 7

Association of enhanced area and transmurality between 2-point bSSFP and LGE. (a) Comparison of the enhanced myocardial area (represented as percent of total myocardial area) using 2-pt bSSFP and LGE revealed a strong association between the two methods (R2 = 0.84) with a slight bias towards larger areas of enhancement with 2-point bSSFP. (b) Bland-Altman plot comparing the difference between enhanced areas by both methods to the mean between both methods revealed a coefficient of variation of 0.204. (c) Similarly, the comparison of the transmurality of enhancement by each method revealed a strong association between 2-point bSSFP and LGE (R2 = 0.73) with (d) Bland-Altman analysis demonstrating a coefficient of variation of 0.0875

Back to article page